Campbell-scientific CURS100 100 Ohm Current Shunt Terminal Input Modul Manuale Utente Pagina 16

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 18
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 15
CURS100 100 Ohm Current Shunt Terminal Input Module
10
BeginProg
Scan (1,Sec,1,0)
'Generic 4-20 mA Input measurement Measure:
VoltDiff (Measure,1,mV2500,1,True,0,_60Hz,0.0625,-25.0)
CallTable (Hourly)
NextScan
EndProg
5.4 CR9000(X) Program Example
VoltDiff(Press_psi, 1, mV5000, 5, 1, 1, 0, 0, 0.3125, 75)
5.5 CR10(X) Program Example
1: Volt (Diff) (P2)
1: 1 Reps
2: 25 ± 2500 mV 60 Hz Rejection Range
3: 1 DIFF Channel
4: 1 Loc [ Press_psi ]
5: 0.3125 Mult
6: 75 Offset
5.6 CR23X Program Example
1: Volt (Diff) (P2)
1: 1 Reps
2: 25 ± 5000 mV, 60 Hz Reject, Fast Range
3: 1 DIFF Channel
4: 1 Loc [ Press_psi ]
5: 0.3125 Mult
6: 75 Offset
Vedere la pagina 15
1 2 ... 11 12 13 14 15 16 17 18

Commenti su questo manuale

Nessun commento