CS10-L and CS15-L Current Transformers
8: Step Loop Index (P90)
1: 2 Step
9: Z=X (P31)
1: 2 X Loc [ BurstAmps ]
2: 4 -- Z Loc [ CS10_1 ]
10: Do (P86)
1: 3 Call Subroutine 3
11: Z=X (P31)
1: 3 X Loc [ Burst_A2 ]
2: 5 -- Z Loc [ CS10_2 ]
12: End (P95)
13: Do (P86)
1: 51 Set Port 1 Low
; This part of the program will store a one minute average of the amperage.
14: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 1 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
15: Set Active Storage Area (P80)^17815
1: 1 Final Storage Area 1
2: 60 Array ID
16: Real Time (P77)^10331
1: 1220 Year,Day,Hour/Minute (midnight = 2400)
17: Average (P71)^5143
1: 64 Reps
2: 4 Loc [ CS10_1 ]
*Table 2 Program
02: 0.0000 Execution Interval (seconds)
*Table 3 Subroutines
;
; Parameter 2 should be 2500 mV for 50-200 amps
; should be 250 mV for 5-49 amps
; should be 25 mV for 0-4.9 amps
; Parameter 5 should be 2.0 msec for 50 Hz or 60 Hz
; Parameter 6 should be 0.05 thousand scans for 50 Hz or 60 Hz
; if parameter 5 & 6 are 2.0 and 0.05, then you have 5 complete cycles at 50 Hz
; or 6 complete cycles at 60 Hz.
1: Beginning of Subroutine (P85)
1: 1 Subroutine 1
12