Appendix C. PT4-L Agitator
C-4
The CR510, CR10X, and CR23X use instruction P86 and P22 as follows:
CR10X P86 and P22 Instructions Example
45: Do (P86)
1: 41 Set Port 1 High
46: Excitation with Delay (P22)
1: 1 Ex Channel
2: 20 Delay W/Ex (units = 0.01 sec)
3: 0 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation
47: Do (P86)
1: 51 Set Port 1 Low
48: End (P95)
The above examples are not as power efficient as possible and would require
AC power to maintain a sufficient battery charge. If it is necessary to operate
an agitator without AC power available, write the program so that the agitator
is only operated for a short period of time just before the measurement is to be
taken.