Campbell-scientific CS547A-L Conductivity/Temperature Probe and A547A Manuale Utente Pagina 21

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 30
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 20
CS547A Conductivity and Temperature Probe and A547 Interface
11: Polynomial (P55)
1: 1 Rep
2: 3 X Loc [T25_01]
3: 4 F(X) Loc [f_of_T ]
4: .99124 C0
5: -1.8817 C1
6: 3.4789 C2
7: -3.51 C3
8: -1.2 C4
9: -43 C5
12: Z=1/X (P42)
1: 4 X Loc [f_of_T ]
2: 6 Z Loc [one_ovrfT ]
13: Z=X*F (P37)
1: 6 X Loc [one_ovrfT ]
2: 1.408 F EC of calibration solution
3: 7 Z Loc [Conductiv]
14: Z=X*Y (P36)
1: 7 X Loc [Conductiv]
2: 1 Y Loc [Rs ]
3: 8 Z Loc [Kc ]
End
7. Maintenance
Routine maintenance includes thoroughly cleaning the orifice of the CS547A
probe with the black nylon brush provided and a little soapy water. Rinse
thoroughly.
8. Analysis of Errors
8.1 EC Measurement Error
1. Bridge Measurement Error: < 1.0%
2. Calibration Error:
bridge measurement: < 0.5%
calibration solution: < 1.0%
3. Ionization Error of KCl and Na+ Solutions After Correction:
< 2.0%, 0.45 to 7.0 mS cm
-1
< 8.0%, 0.005 to 0.45 mS cm
-1
Correction of Ionization Errors: Figures 8.1-1 and 8.1-2 show the amount
of correction applied by the example program to compensate for ionization
effects on the measurements. Also shown is an ideal correction. Factors were
15
Vedere la pagina 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 29 30

Commenti su questo manuale

Nessun commento