Campbell-scientific EC155 CO2 and H2O Closed-Path Gas Analyzer and EC1 Manuale Utente Pagina 57

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 70
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 56
Appendix A. Filter Bandwidth and Time Delay
FIGURE A-2. Frequency response comparison of the EC100 10-Hz
bandwidth and a 50-msec moving average
TABLE A-1. Filter Time Delays for Various Bandwidths
Bandwidth (Hz) Time Delay (ms)
5 800
10 400
12.5 320
20 200
25 160
1 10 70
Hertz
0.0001
0.001
0.01
0.1
1
10
No Units
EC100 10-Hz Bandwidth Filter
10-Hz Bandwidth from a 50-msec Moving Average
EC100 10-Hz Filter Compared to 20-msec Moving Average (Amplitude Responses)
A-3
Vedere la pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 69 70

Commenti su questo manuale

Nessun commento