OPERATOR’S MANUAL OBS500 Smart Turbidity Meter with ClearSensor™ Technology Revision: 5/14 Copyright © 2008- 2014 Campbell

Pagina 2

Table of Contents iv

Pagina 3 - Limited Warranty

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 1. Introduction The OBS500 submersible turbidity meter is designed for general pressure me

Pagina 4 - Assistance

OBS500 Smart Turbidity Meter with ClearSensor™ Technology • The probe must be calibrated with sediments from the waters to be monitored. The procedu

Pagina 5 - Precautions

OBS500 Smart Turbidity Meter with ClearSensor™ Technology • The most important general precaution is to orient the unit so that the OBS sensor looks

Pagina 6

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 4-2. Apply tape to protect sensor FIGURE 4-3. Secure with hose clamps. Do not ov

Pagina 7

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 4-4. Place and secure mounting fixture 5

Pagina 8

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 4.2 Datalogger Programming and Wiring Short Cut is an easy way to program your datalogger t

Pagina 9

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 4. Select Datalogger Model and Scan Interval (default of 5 seconds is OK for most applicati

Pagina 10 - Table of Contents

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 6. After selecting the sensor, click at the left of the screen on Wiring Diagram to see how

Pagina 11 - OBS500 Smart Turbidity Meter

OBS500 Smart Turbidity Meter with ClearSensor™ Technology • Insensitivity to bubbles and organic matter • Ambient-light rejection • Low temperatur

Pagina 13

OBS500 Smart Turbidity Meter with ClearSensor™ Technology Historically, most turbidity sensor manufacturers and sensor users labeled the units NTUs, f

Pagina 14

OBS500 Smart Turbidity Meter with ClearSensor™ Technology The OBS500 has three communication modes: SDI-12, RS-232, or 0 to 5 V. The mode defaults to

Pagina 15

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 6. Specifications Features: • Dual backscatter and sidescatter sensors used to measure tur

Pagina 16

OBS500 Smart Turbidity Meter with ClearSensor™ Technology Diameter: 4.76 cm (1.875 in) Length: 27 cm (10.625 in) Weight: 0.52 kg (1.15 lb) Maximum

Pagina 17

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 7-1. Device Configuration Utility After installing the Device Configuration Utility

Pagina 18 - 5. Overview

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 7-2. Terminal Mode using 1 and H commands TABLE 7-2. RS-232 Terminal Commands Term

Pagina 19 - 5.2 Turbidity Units

OBS500 Smart Turbidity Meter with ClearSensor™ Technology Device Configuration allows you to change the configuration of the OBS500. Select the Setti

Pagina 20 - 5.3 Measurement Details

OBS500 Smart Turbidity Meter with ClearSensor™ Technology TABLE 7-3. OBS500 Connector Pin-Out MCIL-8-MP/MCBH-8-FS Contact Number Electrical Functio

Pagina 21 - to 170°

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 7.3.2 RS-232 Wiring Our CR800, CR850, CR1000, and CR3000 dataloggers have COM ports (contro

Pagina 22 - 6. Specifications

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 7.4 Communication Modes 7.4.1 SDI-12 The OBS500 uses an SDI-12-compatible hardware interfa

Pagina 23 - 7. Installation

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Pagina 24

OBS500 Smart Turbidity Meter with ClearSensor™ Technology TABLE 7-7. SDI-12 and RS-232 Measurement Commands Commands Process Values Returned aM7! a

Pagina 25

OBS500 Smart Turbidity Meter with ClearSensor™ Technology The address may also be changed by connecting to the probe in the Device Configuration Utili

Pagina 26

OBS500 Smart Turbidity Meter with ClearSensor™ Technology For CR10X and CR510 standard mixed-array dataloggers, the datalogger telecommunications comm

Pagina 27 - 7.3.1 SDI-12 Wiring

OBS500 Smart Turbidity Meter with ClearSensor™ Technology universal response curve. However, there is residual nonlinearity that is removed by calibr

Pagina 28 - 7.3.2 RS-232 Wiring

OBS500 Smart Turbidity Meter with ClearSensor™ Technology TABLE 7-9. Calibration Materials and Volumes Side Scatter 90-Degree Materials Calibration

Pagina 29 - 7.4 Communication Modes

OBS500 Smart Turbidity Meter with ClearSensor™ Technology TABLE 7-10. Change in TU value resulting from one hour of evaporation of SDVB standard, i.e

Pagina 30

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 6. Repeat steps 2, 3, 4, and 5 for the other standards. 7. Perform 2nd-order polynomial r

Pagina 31

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 7.5.2.2 Wet-Sediment Calibration Wet-sediment calibration is performed with sediment obtain

Pagina 32 - 7.5 Calibration

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 7-8. Portable Sediment Suspender (left) and OBS beam orientation in suspender

Pagina 33 - Turbidity (NTU)

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 10. Perform 3rd order polynomial regressions on the data to get the coefficients for conver

Pagina 34

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Pagina 35

OBS500 Smart Turbidity Meter with ClearSensor™ Technology Note that Edlog only allocates one input location for Instruction 105. Two input locations

Pagina 36 - 7.5.2 Sediment

OBS500 Smart Turbidity Meter with ClearSensor™ Technology the current usage. Therefore, increased current usage indicates that the wiper needs to be

Pagina 37

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 7-10. Insert screwdriver and rotate clockwise FIGURE 7-11. Shutter disassembled

Pagina 38

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 8. Factors that Affect Turbidity and Suspended-Sediment Measurements This section summarize

Pagina 39 - 7.6 Programming

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 8-2. The apparent change in turbidity resulting from disaggregation methods 8.2 Su

Pagina 40

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 0 20 40 60 80 100 120 140 160 180Scattering Angle0.010.11Relative Scattering IntensityCubesP

Pagina 41

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 8-4. Response of an OBS sensor to a wide range of SSC 8.5 IR Reflectivity—Sediment

Pagina 42

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 8-5. Infrared reflectivity of minerals as a function of 10-Munzell Value 8.6 Wate

Pagina 43 - Sediment Measurements

OBS500 Smart Turbidity Meter with ClearSensor™ Technology more sensitive, by factors of four to six, to mineral particles than particulate organic mat

Pagina 44 - 8.3 Particle-Shape Effects

OBS500 Smart Turbidity Meter with ClearSensor™ Technology FIGURE 9-1. DevConfig, Send OS 10. Troubleshooting A common cause for erroneous, turbidit

Pagina 45 - Scattering Angle

Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO

Pagina 46

OBS500 Smart Turbidity Meter with ClearSensor™ Technology The following three tests are used to diagnose malfunctions of an OBS500. 1. The Finger-Wa

Pagina 47 - 8.7 Bubbles and Plankton

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 11. References Anderson, C.W., 2005, Turbidity (ver. 2.1): U.S. Geological Survey Technique

Pagina 48 - 9. Maintenance

OBS500 Smart Turbidity Meter with ClearSensor™ Technology 42

Pagina 49 - 10. Troubleshooting

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Pagina 50

Appendix A. Importing Short Cut Code 6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section be

Pagina 51 - 11. References

Appendix B. Example Programs B.1 CR1000 SDI-12 Program Although this is a CR1000 program, other CRBasic dataloggers are programmed similarly. &apos

Pagina 52

Appendix B. Example Programs B.2 CR1000 RS-232 Program Although this is a CR1000 program, other CRBasic dataloggers are programmed similarly. '

Pagina 53 - A.1.1 CRBasic Datalogger

Appendix B. Example Programs B.3 CR1000 Analog Program Although this is a CR1000 program, other CRBasic dataloggers are programmed similarly. '

Pagina 54 - A.1.2 Edlog

Appendix B. Example Programs B.4 Examples for High Sediment Loads B.4.1 Normally Open CR1000 Example 'CR1000 Series Datalogger 'OBS500 no

Pagina 55 - B.1 CR1000 SDI-12 Program

Appendix B. Example Programs B.4.2 Cycle Shutter/Wiper for Each Measurement CR1000 Program The following CRBasic program will: • Open the shutter

Pagina 57 - B.3 CR1000 Analog Program

Appendix B. Example Programs B-6

Pagina 58

Appendix C. OBS500 Copper Sleeve Kit Installation 1. Remove the Button Head Hex Screw as shown. 2. Slide the Copper Sleeve over the OBS500 and s

Pagina 59 - Program

Appendix C. OBS500 Copper Sleeve Kit Installation C-2

Pagina 61 - Installation

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Pagina 62

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Pagina 63

Table of Contents 7.6.3 Edlog Programming ...... 29 7.7 Operation in High Sediment Load

Pagina 64

Table of Contents 7-3. Settings Editor screen .................. 16 7-4. Terminal Emulator ..