Model 092 Barometric Pressure Sensor
'Main Program
BeginProg
'Main Scan
Scan(5,Sec,1,0)
'Default Datalogger Battery Voltage measurement 'BattV'
Battery(BattV)
'Default Wiring Panel Temperature measurement 'PTemp_C'
PanelTemp(PTemp_C,_60Hz)
'092 Barometric Pressure Sensor measurement 'BP_mmHg'
VoltSE(BP_mmHg,1,mV5000,1,1,0,_60Hz,0.1,600)
BP_mmHg=BP_mmHg*0.75006
'Call Data Tables and Store Data
CallTable(Table1)
CallTable(Table2)
NextScan
EndProg
8. Maintenance
8.1 Cleaning
Periodic cleaning of the sintered filter at the base of the unit may be required in
dusty conditions. Remove the filter and clean with distilled water.
8.2 Calibration
Calibration is performed against a NIST traceable standard. Calibration
coefficients are stored in non-volatile memory at the factory. No user
calibration procedure is available. The sensor must be returned to Met One
Instruments when calibration is desired. Yearly calibrations are recommended.