Section 7. Installation
262
Table 52. PRTCalc() Type-Code-6 Sensor
Standard ITS-90 SPRT, alpha = 0.003926 (Reference: Minco /
Instrunet)
Constant Coefficient
a 3.9848000E-03
d -2.3480000E-06
e 1.8226630E-05
f -1.1740000E-06
g 1.6319630E+00
h -2.4709290E+00
i 8.8283240E+00
j 2.5091300E+02
7.8.18.2 Measuring PT100s (100-Ohm PRTs)
PT100s (100-ohm PRTs) are readily available. The CR3000 can measure PT100s
in several configurations, each with its own advantages.
The CR3000 is also equipped to measure PRTs with current excitation.
7.8.18.2.1 Self-Heating and Resolution
PRT measurements present a dichotomy. Excitation voltage should be maximized
to maximize the measurement resolution. Conversely, excitation voltage should
be minimized to minimize self-heating of the PRT.
If the voltage drop across the PRT is 50 mV, self-heating should be less than
0.001°C in still air. To maximize measurement resolution, optimize the excitation
voltage (Vx) such that the voltage drop spans, but does not exceed, the ±50-mV
input range.
7.8.18.2.2 PT100 in Four-Wire Half-Bridge
Example Shows:
How to measure a PRT in a four-wire half-bridge configuration
How to compensate for long leads
Advantages:
High accuracy with long leads
Example PRT Specifications:
Alpha = 0.00385 (PRT Type 1)
A four-wire half-bridge, measured with BrHalf4W(), is the best configuration for
accuracy in cases where the PRT is separated from bridge resistors by a lead
length having more than a few thousandths of an ohm resistance. In this example,