Appendix C. 4 to 20 mA Operation
C.1 Setting Operating Parameters using DIP
Switches
The following 4 to 20 mA operating parameters can be set with DIP switches.
DIP switches 1, 2, and 3 are reserved for the SDI-12 interface.
Setting measurement type to level or depth
Level Measurement Set DIP Switch 4 to the OFF position.
Depth Measurement Set DIP Switch 4 to the ON position.
Scaling the measurement
With DIPS 5 and 6 you can scale the available measuring range to a smaller
range. Where the whole measuring range is not required, this has the
advantage of increasing the resolution.
50 ft (not scaled) DIPS 5 and 6 to the ON position.
25 ft DIP 5 to the OFF position, DIP 6 to the ON position.
12 ft DIP 5 to the ON position, DIP 6 to the OFF position.
6 ft DIPS 5 and 6 to the OFF position.
Setting the measurement system
Use DIP 7 to set the engineering units of the measurement.
Metric DIP 7 to OFF.
Imperial DIP 7 to ON.
Setting measurement type
Use DIP 8 to set water level or pressure units.
Water Level DIP 8 to OFF.
Pressure DIP 8 to ON.
C-2