Appendix D. SDI-12 Sensor Support
7. To query the CS470/CS471 for its current SDI-12 address, enter the
command ?!. The CS470/CS471 will respond with the current SDI-12
address.
8. To change the SDI-12 address, enter the command aAb!; where a is the
current address from the above step and b is the new address. The
CS470/CS471 will change its address and the datalogger will exit the SDI-
12 Transparent Mode.
9. Activate the SDI-12 Transparent Mode on Control Port 1 again by entering
1X <enter>. Verify the new SDI-12 address by entering the ?! command.
The CS470/CS471 will respond with the new address.
10. To exit the SDI-12 Transparent Mode, enter *.
FIGURE D-3. CR10X example of using the SDI-12 transparent mode to
change the SDI-12 address from 0 to 1. Sensor is connected to
control port 1.
D-6